Suture fixation of migrated septal occluder device to prevent further migration: a simple surgical technique

Prashant N Mohite12*, Sachin A Kuthe1, Shrirang D Ranade1, Pravin P Kulkarni1, Anton Sabashnikov1 and Aron F Popov23

Author Affiliations

1 Department of Cardiothoracic and Vascular surgery, Postgraduation Institute of Medical Research and Education, Sector- 12, Chandigarh 160012, India

2 Department of Cardiothoracic Transplantation and Mechanical Support, Royal Brompton and Harefield NHS Trust, Hill End Road, UB9 6JH, London, UK

3 Department of Thoracic Cardiovascular Surgery, University of Göttingen, Robert-Koch- Strasse 40, 37099, Göttingen, Germany

For all author emails, please log on.

Journal of Cardiothoracic Surgery 2013, 8:10  doi:10.1186/1749-8090-8-10

The electronic version of this article is the complete one and can be found online at: http://www.cardiothoracicsurgery.org/content/8/1/10


Received:21 August 2012
Accepted:11 January 2013
Published:16 January 2013

© 2013 Mohite et al.; licensee BioMed Central Ltd.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

As the use of percutaneous intervention is increasing for the closure of the atrial septal defect, the procedure related complications are also on rise, migration of the device being most common. The migrated devices with failed percutaneous retrieval must be removed surgically under cardiopulmonary bypass. During establishment of cardiopulmonary bypass, the handling of heart may cause further migration of the device into other chambers of heart which leads to difficulty in finding and retrieval of the device. The authors propose a simple and unique technique to prevent further migration of the septal occluder device.

Keywords:
Atrial septal defect; Septal occlude device; Transcatheter closure; Migration of occlude device

Introduction

The transcatheter closure of the atrial septal defect (ASD) has been proposed as an alternative to surgical therapy for the ostium secundum ASD. The transcatheter closure ASDs was first described by King and Mills in 1976 [1]. It is considered an effective and safe procedure, and although rare but potentially fatal complications have been reported. Migration of the septal occluder devices is a common complication of the procedure [2]. The migrated devices with failed percutaneous retrieval must be removed surgically under cardiopulmonary bypass.

Methods

As soon as the pericardium is opened, the migrated device is palpated. Pre-operative fluoroscopy or simply transthoracic echocardiography locates the migrated device. As the device is palpated, it is held in the pair of forceps along with the wall of the chamber, mostly right atrium. A 3-0 polypropelene suture is entered through the wall of the chamber passing through the migrated device and knot is tied (Figure 1). Aorto- bicaval cannulation is then done and cardiopulmonary bypass is established. Heart arrested with topical hypothermia and high potassium cardioplegia. Right atrium is incised obliquely and the device is found stuck to the atrial wall with the polypropelene suture. The device is retrieved after cutting the suture and atrial septal defect is closed directly or with patch.

thumbnailFigure 1. Suture passing through the occlude device from outside the right atrium.

Discussion

Recently, percutaneous transcatheter procedures emerged as therapeutic alternatives for closure of both atrial septal defects and patent foramen ovale. Percutaneous closure of ASDs is an excellent therapeutic option for the appropriate patient, yielding comparable results to surgery with minimal morbidity and complications. Unfortunately, however, such percutaneous procedures may require surgical intervention for early or late complications. Percutaneous ASD closure has proven to be reliable, safe, and effective [3]. The reported success rate of these devices in closing ASD is as high as 98% [4]. Periprocedural complications include air embolism, device migration, cardiac tamponade, device erosion, atrial arrhythmias, and those related to vascular access [2]. Moreover, device related complications include device migration, device fracture, device thrombosis, and device erosion. Device embolization occurs in about 0.55% of cases, regardless of ASD size, device size, or the physician’s expertise [5].

There are several reasons for the acute failure of these devices. The most important is suboptimal patient and/ or device selection [6,7]. Other suggested mechanisms of acute failure are device related failure, inadequate experience, poor defect rim to hold the device [4-6], and tearing of the interatrial septum as a result of catheter and device manipulations [7]. It has been reported by Balbi M et al. that percutaneous retrieval was successful in 50% of the cases of ASO embolization that have been reported in the literature [8].

Once the device is migrated, it may slip into right atrium, right ventricle to pulmonary artery or into left atrium, left ventricle to aorta. Surgical intervention becomes mandatory when the migrated device could not be retrieved by percutaneous catheter intervention [9]. Urgent surgery and device retrieval is necessary when device gets migrated into heart chambers and causes arrhythmia or hemodynamic compromise.

The migrated septal occluder device may be further migrated due to manipulation of the heart while putting patient on cardiopulmonary bypass. Authors suggest palpating the migrated device gently and holding it with the pair of forceps. The device is then sutured from outside to the wall of the chamber containing it with fine polypropelene suture. Once the device is thus secured to the chamber wall, procedure can be progressed without hesitation and fear of further migration of device. This technique which does not consume much time is useful in occluder device migration in any patient irrespective of age or urgency of the procedure.

Competing interests

The authors declare that they have no competing interests.

Authors’ contribution

PNM: Concept, wrote the manuscript; SDR, PPK and AS: gave important comments; AFP: made critical revision of the article and drafted the manuscript. All authors read and approved the final manuscript.

References

  1. King TD, Thompson SL, Steiner C, Mills NL: Secundum atrial septal defects: nonoperative closure during cardiac catheterization.

    JAMA 1976, 235:2506-2509. PubMed Abstract | Publisher Full Text OpenURL

  2. Post MC, Suttorp MJ, Jaarsma W, et al.: Comparison of outcome and complications using different types of devices for percutaneous closure of a secundum atrial septal defect in adults:a single center experience.

    Catheter Cardiovasc Interv 2006, 67:438-443. PubMed Abstract | Publisher Full Text OpenURL

  3. Butera G, Carminati M, Chessa M, et al.: Percutaneous versus surgical closure of secundum atrial septal defect: comparison of early results and complications.

    Am Heart J 2006, 151:228-234. PubMed Abstract | Publisher Full Text OpenURL

  4. Thanopoulos BD, Laskari CV, Tsaousis GS, et al.: Closure of atrial septal defects with the Amplatzer occlusion device: preliminary results.

    J Am Coll Cardiol 1998, 31:1110-1116. PubMed Abstract | Publisher Full Text OpenURL

  5. Chessa M, Carminati M, Bufera G, Bini RM, Drago M, Rosti L, Giamberti A, Pome G, Bossone E, Frigiola A: Early and late complications associated with transcatheter occlusion of secundum atrial septal defect.

    J Am Coll Cardiol 2002, 39:1061-1065. PubMed Abstract | Publisher Full Text OpenURL

  6. Berger F, Vogel M, Alexi-Meskishvili V, et al.: Comparison of results and complications of surgical and Amplatzer device closure of atrial septal defects.

    J Thorac Cardiovasc Surg 1999, 118:674-680. PubMed Abstract | Publisher Full Text OpenURL

  7. Agarwal SK, Ghosh P, Mittal P: Failure of devices used for closure of atrial septal defects: mechanisms and management.

    J Thorac Cardiovasc Surg 1996, 112:21-26. PubMed Abstract | Publisher Full Text OpenURL

  8. Balbi M, Pongiglione G, Bezante GP: Catheter Cardiovasc Interv.

    Percutaneous rescue of left ventricular embolized amplatzer septal occluder device 2008, 72(4):559-562. OpenURL

  9. Berdat PA, Chatterjee T, Pfammatter JP, Windecker S, Meier B, Carrel T: Surgical management of complications after transcatheter closure of an atrial septal defect or patent foramen ovale.

    J Thorac Cardiovasc Surg 2000, 120:1034-1039. PubMed Abstract | Publisher Full Text OpenURL